BALANCED AMBIENT CALORIMETER

Climatic Testing Systems, Inc. (CTS) Balanced Ambient Calorimeters are designed in accordance with Ashrae Standard 16 and are used by air conditioner manufacturers to determine the rated capacity of window and split units.  These systems can be supplied as calibrated room type or full balanced ambient with a conditioned annular space around the test room.  Room temperature and humidity conditions are precisely and accurately controlled.  All power inputs to the inner rooms and the test unit are measured and balanced.  The calorimeters are provided with a fully automatic, computerized data acquisition and control system.

 
Typical Performance Specifications
Parameter Range Stability

Indoor Room
Dry Bulb Temperature 60 to 95°F ± 0.15°F
Dew Point Range 50 to 68°F ± 0.15°F
Uniformity   ± 0.5°F

Indoor Room Annular Space
Dry Bulb Temperature 60 to 95°F ± 0.5°F
Dew Point Range 50 to 68°F ± 1.0°F
Uniformity   ± 1.0°F

Outdoor Room
Dry Bulb Temperature 45 to 120°F ± 0.15°F
Dew Point Range 38 to 65°F ± 0.15°F
Uniformity   ± 0.5°F

Outdoor Room Annular Space (Optional)

Dry Bulb Temperature 45 to 120°F ± 0.5°F
Dew Point Range 38 to 65°F ± 0.5°F
Uniformity   ± 1.0°F

 
  > Capacity Range: 6000 to 36,000 BTU/hr
     
  Software Features
  > Editing of test program
  > Automatic test sequencing
  > Loop calibration and calibration reports
  > Hardware checkout routine
  > Stability monitoring
  > Real-time tolerance adjustments
  > Real-time history graphing capability
  > Automatic test report generation
  > Data file management
 
     
  Performance Measurement
  > 2% agreement or better when compared to a calibrated unit
  > 1% repeatability of test results
 
     
  Options
  > Automated pressure equalizer
  > Air flow code tester for airflow measurement
  > Outdoor room controlled annular space
  > Heating capacity test capability without reversing UUT mounting
   
  The specifications shown above are typical. CTS can customize your system to provide the performance that you desire.
   